P374 – UHV sputter deposition cluster tool TU/e

Application

UHV sputter deposition cluster tool for thin film and multilayer deposition at 2″ substrates

Year of delivery

2014

Installation site

Technical University of Eindhoven, Netherland

Design Features

  • UHV multichamber magnetron sputter deposition system with combination of confocal and face to face sputtering.
  • The standard sample size and shape is 2″ wafer. Special shaped RHK STM samples possible too.
  • UHV deposition chamber with nine 2″ magnetrons, three 3″ magnetrons and one multipocket e-beam source installed.
    • Except of two 2″ magnetrons all sources can be used in face to face as well as confocal configuration.
    • All magnetrons with manual in situ tilting and easy change able magent system.
    • Linear multipocket e-beam with manual pocket exchange system and thickness controller incl. rate regulation & end point detection.
    • Fully motorized 3 axes sample manipulator with integrated motorized wedge shutter, RF bias potential and maximal sample temperature above 850°C.
    • Residual gas analyser system.
  • UHV sample preparation chamber with gridded DC ion source, DC plasma oxidation setup and magnet stage installed.
    • Fully motorized 2 axes sample manipulator with integrated DC bias potential and maximal sample temperature well above 900°C.
    • Manual translating plasma oxidation stage.
    • Magnet stage with motorized translation axis.
  • UHV surface analytik chamber with XPS, UPS, ion source, electron source, flood gun and ErLEED installed.
    • Fully manual 4 axes sample manipulator with integrated DC bias potential and maximal sample temperature above 900°C.
    • Manual translating UV source.
    • Manual translating X-Ray source with monochromator.
    • Manual translating X-Ray source without monochromator.
  • Storage chamber for 16 samples as part of the linear transfer system.
  • Load lock chamber with storage for 6 samples, installed lamp heating option and connection port to a mini glove box.
  • Semiautomatic sample transfer between all chambers via a linear distribution chamber system.
  • One STM chamber system provided by RHK is connected to the linear transfer chamber system.
  • Integrated bake out system with bake out tents at all chambers (except of load lock chamber).

Special Features

  • Titanium sublimation pumps (TSP) with cryo shroud installed at all chambers (except of load lock).
  • Handling incl. in situ exchange of sample masks in the sputter chamber.
  • Different sample sizes and shapes from 2″ wafer down to 10mm x 10mm samples can be handled (using different kind of sample holders).
  • Thin film deposition at STM tips possible in sputter chamber using special sample holder with included tip holder.
  • Parts of the system are placed in a second lab room (linear transfer moves through a hole in the wall between both rooms).

Outer Dimensions

Technical specifications and performance values

General

Sputtering chamber

Size

1100 mm diameter, about 1160 mm height

Material

stainless steel

Sample preparation chamber

Size

300 mm diameter, about 930 mm height

Material

stainless steel

Analytik chamber

Size

240 mm diameter, about 880 mm length

Material

stainless steel

Load lock chamber

Size

160 mm diameter, about 300 mm height

Material

stainless steel

Linear transfer chamber

Size

4 segments with 165 mm diameter, about 2040 mm length each

Material

stainless steel

Vacuum

Sputtering chamber

Base pressure

< 2 * 10-10 mbar

Pump down time

2 hours to < 10-7 mbar

Chamber pumping

Turbo pumping stage in combination with titanium sublimation pumping stage, chamber lid as well as sample manipulator differentially pumped by dry foreline pump

Bake out

< 180°C

Sample preparation chamber

Base pressure

< 2 * 10-10 mbar

Pump down time

3 hours to < 10-7 mbar

Bake out

< 180°C

Chamber pumping

Turbo pumping stage in combination with titanium sublimation pumping stage and dry foreline pump

Analytik chamber

Base pressure

< 10-10 mbar

Pump down time

2 hours to < 10-7 mbar

Bake out

< 180°C

Chamber pumping

Turbo pumping stage in combination with titanium sublimation pumping stage and dry foreline pump

Load lock chamber

Base pressure

< 10-8 mbar

Pump down time

1/2 hour to < 10-6 mbar

Chamber pumping

Turbo pumping stage with dry foreline pump

Linear transfer chamber

Base pressure

< 2 * 10-10 mbar

Pump down time

< 3 hours to < 10-7 mbar

Chamber pumping

Ion getter pumping stage in combination with titanium sublimation pumping stage

Bake out

< 180°C

Manipulator features

Sputtering chamber

Sample size

diameter max. 2″ substrate

Motion axes

3 motorized axes (manipulator arm rotation, z tranlsation and (continous) sample stage rotation)

Motorized wedge shutter (part of the manipulator head) with a motion speed of min. about 0.4µm/s and max. about 5mm/s, incl. speed profile feature

Temperatures

Room temperature (not stabilized) up to 830°C at sample

Special features

RF bias potential option

Handling of sample masks

Handling of sample holder for STM tip coating

Sample preparation chamber

Sample size

diameter max. 2″ substrate

Motion axes

2 motorized axes (manipulator z translation and (continous) rotation of the sample stage)

Temperatures

Room temperature (not stabilized) up to 990°C at sample

Special features

DC bias potential option

Analytik chamber

Sample size

diameter max. 2″ substrate

Motion axes

4 manual axes (manipulator x, y & z translation and sample stage tilting)

Temperatures

Room temperature (not stabilized) up to 900°C at sample

Special features

DC bias potential option

Load lock chamber

Sample size

diameter max. 2″ substrate (max. 6 pieces)

Motion axes

2 manual axes (rotation, z translation)
Rotation axis equipped with an air side indexer plate for easy and fast sample loading via access door or transfer rod

Temperatures

Room temperature (not stabilized) up to 300°C at sample (not stabilized)

Sample preparation features

Sample preparation chamber

Oxydation /

Plasma treatment

max. 0.1 mbar partly ionised oxygen gas (using a TU/e designed oxidation plasma setup)

Gas mixture variable from pure argon up to almost pure oxygen

Ion beam etching /

sample precleaning

Wide range variation of ion source to sample distance

Wide range variation of ion energy and ion beam current

Performance test results

Long time heating (SP chamber)
RGA at base pressure (SP chamber)
Long time heating (Sample preparation chamber)
Magnet performance (Sample preparation chamber)
Long time heating (Analytic chamber)
Long time heating (Load lock chamber)