Loading ...
P399 - Exit slit unit for soft X-ray synchrotron radiation at XM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH
P399 - Exit slit unit for soft X-ray synchrotron radiation at XM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH

P399 – Exit Slit Unit XM BESSY/HZB

Application

Exit slit unit for soft X-ray synchrotron radiation

Year of delivery

2015

Installation site

XM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany

Design Features

  • Double slit mechanism – vertically and horizontally defining slits.
  • Motorized and encoded slits opening, symmetric opening, non-linear opening characteristics.

  • Hardened stainless steel slit blades.
  • Electrical isolation and gold coating of the horizontally defining slit blades.

  • Motorized and encoded slit unit translation along beam axis.
  • Motorized and encoded translation of the horizontally defining slit transversal to the beam axis.
  • Manual rotation of the vertically defining slit.
  • Bias plate coated with fluorescent material.
  • Adjustable support structure.

Performance Features

  • Precision and slit opening repeatability smaller than 100 nm.
  • Parallelism smaller than 0.8 µm over 10 mm edge length.
  • Straightness smaller than 0.5 µm over 10 mm edgle length.
  • Very small deviation (submicrometer) of the slit opening calibration curve from the theoretical non-linear slit opening function.

Outer Dimensions

P399 - Outer dimensions - Exit slit unit for soft X-ray synchrotron radiation at XM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH

Technical specifications and performance values

General

Vacuum chamber

Chamber type

cylindrical, horizontal axis

Main flanges

DN160CF cluster flanges

Material

1.4301

Base pressure

10-10 mbar range

Translation along beam axis

Range

400 mm

Resolution (design value)

0.008 µm /motor fine step
(1/128, recommended)
0.01 µm /encoder count

Support structure
adjustment ranges

Longitudinal

± 10 mm

Vertical

± 10 mm

Lateral

± 10 mm

Horizontally defining slit (HDS)

Slit blades

Material

stainless steel 1.7218 with hardened knife edge and gold coating

Straightness

< 0.15 µm over 10 mm (1σ)

Parallelism

0.01 µm /mm
(10 µm slit opening)
0.006 µm /mm
(25 µm slit opening)
0.03 µm /mm
(50 µm slit opening)

Slit opening

Range

5.8 µm to 8 mm

Repeatability

± 0.03 µm
(up to 350 µm slit opening)

Resolution (design value)

0.003 µm /motor fine step
(1/32, recommended)
0.01 µm /encoder count

Slit translation
transversal to the beam axis

Range

8 mm

Resolution (design value)

0.003 µm /motor fine step
(1/32, recommended)
0.01 µm /encoder count

Vertically defining slit (VDS)

Slit blades

Material

stainless steel 1.7218 with hardened knife edge

Straightness

< 0.25 µm over 10 mm (1σ)

Parallelism

0.07 µm /mm
(10 µm slit opening)
0.04 µm /mm
(25 µm slit opening)
0.02 µm /mm
(50 µm slit opening)

Slit opening

Range

5.5 µm to 9.7 mm

Repeatability

± 0.05 µm
(up to 350 µm slit opening)

Resolution (design value)

0.003 µm /motor fine step
(1/32, recommended)
0.01 µm /encoder count

Slit rotation

Range

± 1° (in air )
± 0.5° (in UHV)

Slit width variation upon slit rotation

0.77 µm (1σ)

Performance Test Results

Slit width calibration (HDS)
P399 - Slit width calibration (HDS) - Exit slit unit for soft X-ray synchrotron radiation at XM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH
Slit opening repeatability (HDS)
P399 - Slit opening repeatability (HDS) - Exit slit unit for soft X-ray synchrotron radiation at XM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH
Slit blades parallelism (HDS)
P399 - Slit blades parallelism (HDS) - Exit slit unit for soft X-ray synchrotron radiation at XM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH
Slit blades parallelism (HDS)
P399 - Slit blades parallelism (HDS) - Exit slit unit for soft X-ray synchrotron radiation at XM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH
Slit blades parallelism (HDS)
P399 - Slit blades parallelism (HDS) - Exit slit unit for soft X-ray synchrotron radiation at XM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH
Slit width calibration (VDS)
P399 - Slit width calibration (VDS) - Exit slit unit for soft X-ray synchrotron radiation at XM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH
Slit opening repeatability (VDS)
P399 - Slit opening repeatability (VDS) - Exit slit unit for soft X-ray synchrotron radiation at XM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH
Slit blades parallelism (VDS)
P399 - Slit blades parallelism (VDS) - Exit slit unit for soft X-ray synchrotron radiation at XM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH
Slit blades parallelism (VDS)
P399 - Slit blades parallelism (VDS) - Exit slit unit for soft X-ray synchrotron radiation at XM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH
Slit blades parallelism (VDS)
P399 - Slit blades parallelism (VDS) - Exit slit unit for soft X-ray synchrotron radiation at XM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH
Slit width variation upon slit rotation (VDS)
P399 - Slit width variation upon slit rotation (VDS) - Exit slit unit for soft X-ray synchrotron radiation at XM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany - Bestec GmbH