Loading ...
P354 - Primary slits for soft X-ray synchrotron radiation at Beamline I05-ARPES at Diamond Light Source, Didcot, United Kingdom - Bestec GmbH
P354 - Primary slits for soft X-ray synchrotron radiation at Beamline I05-ARPES at Diamond Light Source, Didcot, United Kingdom - Bestec GmbH

P354 – Primary Slits I05-ARPES DLS

Application

Primary slits for soft X-ray synchrotron radiation

Year of delivery

2012

Installation site

Beamline I05-ARPES at Diamond Light Source, Didcot, United Kingdom

Design Features

  • Vertically and horizontally defining slit blades.
  • Motorized and encoded opening; independent translation of the slit blades.
  • Tungsten slit blades.
  • Actively cooled OFHC absorber blocks attached to all slit blades.

  • Temperature sensor available at each blade.
  • Adjustable support structure.

Performance Features

  • Power density: 60 W /mm2.
  • Total absorbed power: 2 kW in 13 x 13 mm2 area.

Outer Dimensions

P354 - Outer dimensions - Primary slits for soft X-ray synchrotron radiation at Beamline I05-ARPES at Diamond Light Source, Didcot, United Kingdom - Bestec GmbH

Technical specifications and performance values

General

Vacuum chamber

Chamber type

cylindrical, horizontal axis

Material

1.4301

Base pressure

10-9 mbar range

Primary slits

Slit blades

Material

Tungsten (water cooled)

Straightness

< 0.10 µm over 10 mm

Parallelism

< 5 µm over 10 mm

Slit blades translation

Range

-5 mm to +10 mm

Repeatability

± 0.20 µm

Resolution (design value)

0.006 µm /motor fine step
(1/16, recommended)
0.05 µm /encoder count

Performance Test Results

Slit blades translation repeatability
P354 - Slit blades translation repeatability - Primary slits for soft X-ray synchrotron radiation at Beamline I05-ARPES at Diamond Light Source, Didcot, United Kingdom - Bestec GmbH
Slit blades straightness
P354 - Slit blades straightness - Primary slits for soft X-ray synchrotron radiation at Beamline I05-ARPES at Diamond Light Source, Didcot, United Kingdom - Bestec GmbH
Slit blades straightness
P354 - Slit blades straightness - Primary slits for soft X-ray synchrotron radiation at Beamline I05-ARPES at Diamond Light Source, Didcot, United Kingdom - Bestec GmbH